- Ranges 0 to 1 psi through to 10,000 psi
(0 to 70 mbar through to 700 bar)
- 0.1% full scale (FS) accuracy
- Dimensions: 190 x 90 x 36mm.
- Weight: 500g.
- Rugged, lightweight handheld design
- Leak test, tare, max/min and filter
- Intrinsically safe version
- Standard ranges available from stock
- Optional pneumatic and hydraulic hand pumps
DPI 705 internal sensor version
The DPI 705 includes an internal sensor and has a
1/8 in NPT female or G 1/8 female (with 4 mm ID hose
adaptor) pressure connector. Most internal ranges are
available from stock.
DPI 705R remote sensor version
The remote sensor of the DPI 705R has an integral cable
connector, a 3 ft (1 m) mating cable and a 1/4 in NPT
female or G 1/8 female pressure connector.
Accuracy
±0.1% FS non-linearity, hysteresis and repeatability.
Environmental
Temperature Performance Averaged
32°F to 122°F (0°C to 50°C)
- ±0.01% rdg/°F (±0.02% reading/°C)
- For absolute ranges < = 15 psi (1 bar) add
±0.025% FS/°F (±0.05% FS/°C)
- For absolute ranges > 15 psi (1 bar) add
±0.01% FS/°F (±0.02% FS/°C)
- Calibrated range: 32°F to 122°F (0°C to 50°C)
- Operating range: 15°F to 122°F (-10°C to 50°C)
Humidity
0 to 95% non condensing.
Overpressure
Safe to 2 x FS. Display flashes at 110% FS.
Media Compatibility
- Internal sensor: stainless steel, except positive
differential non-corrosive fluid and negative differential
dry gas.
- Remote sensor: stainless steel, except negative
differential non-corrosive fluid.
Safety Conformance
EN61010, EN50081-1, EN50082-2, CE marked
Intrinsically safe version
- Ex ia IIC T4 (-10°C to 50°C)
to CAN/CSA E79-11-95 and CAN/CSA E79-0-95,
(Class I, Division 1, Groups A,B,C&D)
- EEx ia IIC T4 (-10°C to 50°C)
to ATEX EN50020:1994 and EN500014:1992
- Cena:
This e-mail address is being protected from spambots. You need JavaScript enabled to view it.
-