Performance | |
Measurement Technology | Thermo-paramagnetic oxygen sensor |
Gas | Process and non-condensing sample with particles <3µm |
Measurement Range | Selectable from 0–0.5% up to 0-50% and 20, 80 or 90–100% |
Display Resolution | 0.01% |
Display Type | Backlit LCD |
Accuracy Excluding Suppressed Zero Ranges | <±1% of span or ±0.02% O2 whichever is greater |
Accuracy For Suppressed Zero Ranges (80/90 - 100%) | < 0.2% O2 |
Response Time (T90) with High Speed Response (HSR) Enabled | < 15 seconds |
Repeatability | ±0.2% of span or 0.01% O2 whichever is greater |
Linearity | ±0.5% of span or 0.05% O2 whichever is greater |
Zero Stability | ±0.25% of span per month |
Span Stability | ±0.25% of span per month |
Sample Flow Rate | 100 to 600 ml/min |
Sample Flow Effect (calibrated at 300 ml/min) | < 1% of span for flows: 100 to 600 ml/min |
Sample Pressure | 0 to 3 barg inlet with atmospheric vent |
Sample Temperature | 0 to +55°C max (+45°C for standard cell) |
Sample Cell Temperature | Standard +50°C (Optional +55°C and +60°C) |
Background Gas | Unit can be calibrated in a user-defined background gas to increase the accuracy |
Electrical Specifications | |
Analog Inputs | 2 off 4-20 mA inputs One for an external sensor that can be displayed on the screen One to act as an active compensation for the process conditions |
Analog Outputs | 2 off 4-20 mA outputs (isolated) |
Output Ranges | Primary range is set to the calibrated range of the instrument The second is user selectable within the primary range |
Alarms | 2 off single pole changeover (SPCO) relays for O2 concentration (250 V, 5 A max) |
Datalogging | The user can use the digital communications to log the output from the analyzer The unit will store 40 alarm points and the min/max O2 concentrations with date and time stamp |
Digital Communications | MODBUS RTU over RS485 Protocol |
Power Supply | 24 V DC; 1.5 A max |
Operating Conditions | |
Ambient Temperature | Dependent on o-ring selection: Silicon: –40°C to +55°C Viton®: –15°C to +55°C Ekraz: –10°C to +55°C |
Mechanical Specifications | |
Warm Up Time | < 30 minutes |
Stabilization Time | 15 minutes |
Dimensions | 234 x 234 x 172mm (w x d x h) |
Weight | 9.7kg |
Wetted Materials | 316 & 430F stainless steel, boroscillicate glass, platinum, 3M 2216, plus O-ring |
O-Ring Materials | Viton, Silicone or Ekraz |
Gas Connection | 1/4” NPT, 1/4” tube or 6mm tube |
Ingress Protection | IP66, NEMA 4X |
Hazardous Area Classification | |
ATEX | II 2GD Ex d IIB +H2 T3 Gb Ex tb IIIC T137°C Db IP66 |
IECEx | Ex d IIB +H2 T3 Gb Ex tb IIIC T137°C Db IP66 |
Temperature Ranges for ATEX and IECEx as per O-ring Type | Silicon: Ta = -40°C to +55°C Viton®: Ta = -15°C to +55°C Ekraz: Ta = -10°C to +55°C |
CCSAUS | Class I, Division 1, Groups B,C,D |
INMETRO | Ex d IIB + H2 T3 Gb Ex tb IIIC T137°C Db IP66 |
TC TR EX | 1 Ex d IIB + H2 T3 Gb |
The XTP601 process oxygen analyzer meets or exceeds all relevant clauses in BS EN 50104 : 2010 “Electrical apparatus for the detection and measurement of oxygen”. |
Pirkt XTP601 - oxygen analyzer for safe or hazardous areas
Ja Jums ir kādi jautājumi par šo produktu, nekautrējieties, lūdzu sazināties ar mums pa
e-pastu: This email address is being protected from spambots. You need JavaScript enabled to view it. vai pa tālruni: +370 640 20 911.
XTP601 - oxygen analyzer for safe or hazardous areas, cena: --- euro
Parakstoties adresātu jaunumiem, jums tiks sūtītas jaunākās ziņas no mums.
SIA "Tek Know Baltic" Latvija, Riga, Ūdens iela 23, LV-1007
Tel: +370 640 20 911, E-pasts: info@tekknow.lv
© 2018, SIA "Tek Know Baltic". Visas tiesības aizsargātas.