Main Unit Measurement Parameters | |
Channel Configuration | Single and dual channel |
Moisture Content | ppmw |
Temperature | °C and °F |
Water Dew Point | °C and °F |
Sample Flow Rate | Status indication: Low-Good-High (<0.1 to >0.3 l/min) |
Sensor Technology | |
Moisture Sensor Technology | Michell Ceramic Moisture Sensor |
Calibration Range | -100 to +20°Cdp |
Measurement Range | -120 to +30°Cdp (0.001 ppmw to saturation) |
Accuracy | Dew point: ±1°C between -59.9 & +20°Cdp Moisture content: ±10% of reading Dew point: ±2°C between -60 & -100°Cdp Moisture content: ±20% of reading Analysis temperature: Accuracy ±0.2°C |
Resolution | 0.1°C between +20 and -80°Cdp 1°C between -80 and -100°Cdp |
Temperature coefficient | Algorithm compensation |
Operating Pressure | Max. 50 barg |
Sample Flow | Up to 1 l/min (0.1 - 0.2 l/min recommended) |
HMI | |
Keypad/Interface | Capacitive touch-screen through glass |
Display | Vacuum fluorescent |
Datalogging | A rolling maximum of 150 data logs is available Each log records time, date, moisture and temperature values for each channel 5 minutes minimum and 60 minutes maximum logging intervals can be set by the user |
Communications and Output | Two non-isolated 4-20 mA per measurement channel 500 Ω maximum load Range and parameter settable by user Modbus RTU @ 9600 baud-rate Alarms: two volt free contacts per channel; one process value and one instrument status Also available via Modbus communications |
Enclosure | |
Type | Flameproof EExd |
Construction | Cast copper-free aluminium (IP66/NEMA 4) |
Finish | Chromate primer, Polyester P9010 powder coated (black). Meets BS3900 |
General | |
Sample Connections | 1/4" NPT (female) |
Weight | 22.5kg |
Operating Supply Voltage | 90 to 260 V AC, 50/60 Hz, 180 W Main Unit |
Operating Environment | -20 to +60°C up to 95% RH Indoor or Outdoor, shaded location |
Certification | |
Hazardous Area Certification | ATEX Certification: II 2 G Code: EEx d IIB + H2 T4 (135°C) IP66 Tamb -40 to +60°C CCSAUS approved: Div 1, Class 1, Group B,C,D GOST Ex |
Pattern Approval | GOST-R, GOST-K |
Remote Interface Optional for indoor safe area installation | |
Host Capacity | Up to 31 Michell EExd Process Area Installation Instrumentation Range analysers in any combination of models via Modbus |
Display | Backlit color 15cm LCD |
User Interface | Integral mouse, on-screen keyboard |
Function | View and modify analyser operating settings Record parameters to file and view virtual chart recorder analyser diagnostics |
Ports | LAN, USB2, Modbus Re-transmission |
Data Download | CSV (comma separated variable) format for easy import to ExcelTM |
Remote Access | OPC (OLE for Process Control) over a LAN Embedded web page/unique IP address for PC access Remote data reading via Modbus RTU |
Power Supply | 100-280V, 50/60Hz |
Plant control integration | ActiveX components to be used by software programmers for integration of Modbus protocol into a general plant software system |
If you have any questions about this product feel free to contact with us by