General specifications | |
Power requirements | 100 V ac to 240 V ac, 47 Hz to 63 Hz |
Fuse | T2A 250 V ac |
Max power consumption | 100 W |
Operating ambient temperature range | 15 °C to 35 °C |
Storage temperature | -20 °C to 70 °C |
Relative humidity | Operating: <80 % to 30 °C, <70 % to 40 °C, <40 % to 50 °C |
Storage: <95 %, non-condensing. A power stabilization period of four days may be required after extended storage at high temperature and humidity. | |
Vibration | MIL-T-28800 |
Altitude (operation) | <2000 m |
Ingress protection | IEC 60529: IP20 |
Safety | IEC 61010-1, Installation Category II, Pollution degree 2 |
Weight (Chassis only) | 13 kg (28.5 lbs) |
Dimensions | Height: 147 mm (5.78 in) |
Width: 454 mm (17.79 in) | |
Depth 488 mm (19.2 in) | |
Rack mount dimensions | 3U-19 inch rack |
Warm up time | 15 minutes typical |
Control specifications | |
Control precision (dynamic mode) | PM200-BG2.5K range ± 0.005% range All other ranges ± 0.001% range |
Control turndown* | 10:1 (typical) |
Low control point | 1 kPa (0.15 psi) absolute |
Supply pressure requirements | |
Clean dry N2 or air – Industrial grade nitrogen, 99.5 %+ | |
Particulate contamination | ≤ 1.25 micrometer (50 microinches) |
Maximum moisture content | -50 °C dew point |
Maximum hydrocarbon content | 30 ppm |
Vacuum supply requirements (when operating near or sub-atmospheric) | |
>50 liters per minute capacity with Auto Vent feature | |
Appropriate protections for High Pressure Gauge work system exhaust gas will pass through the Vacuum supply system. | |
Interface/communications | |
Primary remote interfaces | IEEE, Ethernet, RS232, USB |
System connection | Supports interconnection of two or three systems |
Switch test connection | Standard 4 mm jack: Nominal 24 v dc isolated drive Maximum 30 V dc w.r.t. chassis ground |
Aux drivers | 4 external solenoid drivers 24 V dc drive (maximum drive 6 W continuous per channel) |
* Control turndown is defined as the relationship between the provided supply pressure and the appropriate supply pressure for the range. For example, a unit with a 7 MPa (1000 psi) and 700 kPa range (100 psi) with a supply pressure of 7.7 MPa (1100 psi) will provide control precision of 0.001% range because 7 MPa is 10 times greater than 700 kPa. A system with ranges of 20 MPa (3000 psi) and 700 kPa (100 psi) with supply pressure of 22 MPa (3300 psi) will have 0.001% range control precision on the 20 MPa range but only 0.003% control precision on the 700 kPa range. Control precision of 0.001% on the low range can be achieved by reducing the supply pressure. |
Model | Gauge mode range (SI units) | Absolute mode range (SI units) |
Gauge mode range (Imperial units) | Absolute mode range (Imperial units) | 1 year Specification | ||
Relative uncer- tainty (% Reading) |
Threshold uncer-tainty (% Span) |
Absolute mode adder (% Full scale) |
|||||
PM600-BG15K | -15 to 15 kPa | - | -60 to 60 inH₂0 | - | 0.01 % | 0.003 % | - |
PM600-G100K | 0 to 100 kPa | - | 0 to 15 psi | - | 0.01 % | 0.003 % | - |
PM600-G200K | 0 to 200 kPa | - | 0 to 30 psi | - | 0.01 % | 0.003 % | - |
PM600-A100K | -100 to 0 kPa | 6 to 100 kPa | -13.8 to 0 psi | 0.9 to 15 psi | 0.01 % | 0.003 % | 0.007 % |
PM600-A200K | -90 to 100 kPa | 10 to 200 kPa | -13.2 to 15 psi | 1.5 to 30 psi | 0.01 % | 0.003 % | 0.007 % |
PM600-A350K | -90 to 250 kPa | 10 to 350 kPa | -13.2 to 35 psi | 1.5 to 50 psi | 0.01 % | 0.003 % | 0.007 % |
PM600-A700K | -82 to 700 kPa | 18 to 700 kPa | -12.1 to 100 psi | 2.6 to 100 psi | 0.01 % | 0.003 % | 0.007 % |
PM600-A1.4M | -0.065 to 1.4 MPa | 0.035 to 1.4 MPa | -10 to 200 psi | 5 to 200 psi | 0.01 % | 0.003 % | 0.007 % |
PM600-A2M | -0.03 to 2 MPa | 0.07 to 2 MPa | -5 to 300 psi | 10 to 300 psi | 0.01 % | 0.003 % | 0.007 % |
PM600-A3.5M | -0.03 to 3.5 MPa | 0.07 to 3.5 MPa | -5 to 500 psi | 10 to 500 psi | 0.01 % | 0.003 % | 0.007 % |
PM600-A7M | 0 to 7 MPa | atmosphere to 7 MPa | 0 to 1000 psi | atmosphere to 1000 psi | 0.01 % | 0.003 % | 0.007 % |
PM600-A10M | 0 to 10 MPa | atmosphere to 10 MPa | 0 to 1500 psi | atmosphere to 1500 psi | 0.01 % | 0.003 % | 0.007 % |
PM600-A14M | 0 to 14 MPa | atmosphere to 14 MPa | 0 to 2000 psi | atmosphere to 2000 psi | 0.01 % | 0.003 % | 0.007 % |
PM600-A20M | 0 to 20 MPa | atmosphere to 20 MPa | 0 to 3000 psi | atmosphere to 3000 psi | 0.01 % | 0.003 % | 0.007 % |
BRM600-BA100K | - | 70 to 110 kPa | - | 10 to 16 psi | 0.01 % | - | - |
* 1-year specification is instrumental measurement uncertainty that includes linearity, hysteresis, resolution, reference uncertainty, 1-year stability, and temperature effects from 15 to 35 °C. The uncertainties are provided at 95% confidence, k=2, normally distributed. | |||||||
** Gauge mode uncertainty is the greater of the relative uncertainty and the threshold uncertainty. For absolute ranges used in gauge mode there is an additional uncertainty of ± 7 Pa for dynamic barometric compensation. When combined with other uncertainties this changes the threshold uncertainty for the PM600-A100K to 0.008% Span and for the PM600-A200K to 0.004% Span. | |||||||
** Absolute mode uncertainty is the greater of the relative uncertainty and the threshold uncertainty plus the absolute mode uncertainty adder for the lowest range PM600-AXXX module installed. For example, if a PM600-A200K and a PM600-A2M, the uncertainty at 2000 kPa absolute would be 0.2 kPa (0.01% * 2000 kPa) plus 0.014 kPa. |
Model |
Range (SI units) |
Range (Imperial units) |
Measurement mode |
Uncertainty (%FS) |
PM200-BG2.5K |
-2.5 to 2.5 kPa |
-10 to 10 inH₂0 |
gauge |
0.20 % |
PM200-BG35K |
-35 to 35 kPa |
-5 to 5 psi |
gauge |
0.05 % |
PM200-BG40K |
-40 to 40 kPa |
-6 to 6 psi |
gauge |
0.05 % |
PM200-A100K |
2 to 100 kPa |
0.3 to 15 psi |
absolute |
0.10 % |
PM200-BG100K |
-100 to 100 kPa |
-15 to 15 psi |
gauge |
0.02 % |
PM200-A200K |
2 to 200 kPa |
0.3 to 30 psi |
absolute |
0.10 % |
PM200-BG200K |
-100 to 200 kPa |
-15 to 30 psi |
gauge |
0.02 % |
PM200-BG250K |
-100 to 250 kPa |
-15 to 36 psi |
gauge |
0.02 % |
PM200-G400K |
0 to 400 kPa |
0 to 60 psi |
gauge |
0.02 % |
PM200-G700K |
0 to 700 kPa |
0 to 100 psi |
gauge |
0.02 % |
PM200-G1M |
0 to 1 MPa |
0 to 150 psi |
gauge |
0.02 % |
PM200-G1.4M |
0 to 1.4 MPa |
0 to 200 psi |
gauge |
0.02 % |
PM200-G2M |
0 to 2 MPa |
0 to 300 psi |
gauge |
0.02 % |
PM200-G2.5M |
0 to 2.5 MPa |
0 to 360 psi |
gauge |
0.02 % |
PM200-G3.5M |
0 to 3.5 MPa |
0 to 500 psi |
gauge |
0.02 % |
PM200-G4M |
0 to 4 MPa |
0 to 580 psi |
gauge |
0.02 % |
PM200-G7M |
0 to 7 MPa |
0 to 1000 psi |
gauge |
0.02 % |
PM200-G10M |
0 to 10 MPa |
0 to 1500 psi |
gauge |
0.02 % |
PM200-G14M |
0 to 14 MPa |
0 to 2000 psi |
gauge |
0.02 % |
PM200-G20M |
0 to 20 MPa |
0 to 3000 psi |
gauge |
0.02 % |
* Gauge mode modules (PM200-GXXX or PM200-BGXXX) with ranges of 100 kPa (15 psi) or greater will support absolute mode measurement when used with a Barometric Reference Module. |
Pirkt Fluke 6270A - modulārs spiediena kontrolieris
Ja Jums ir kādi jautājumi par šo produktu, nekautrējieties, lūdzu sazināties ar mums pa
e-pastu: This email address is being protected from spambots. You need JavaScript enabled to view it. vai pa tālruni: +370 640 20 911.